Rapidlab 1200 Operator’s Guide: Operating the System 2-27
02087462 Rev. V
Using Patient Results
This section provides information about viewing and recalling patient results.
Viewing Patient Results
After you complete the Data Entry screen, the Results screen displays.
• The screen displays result values in yellow while analysis is in progress.
Preliminary results are not displayed if the system is performing additional
calibrations during analysis.
• The Results screen displays a parameter as gray if a result cannot be reported
because of system problems, such as a parameter that failed QC.
• The Results screen does not display a parameter if you did not enter a required
value, such as the temperature for a temperature corrected value.
• The Results screen displays the following symbols in red.
These symbols display with the results on the screen and in the printed report:
CAUTION: Only use final results for diagnostic purposes. Never use preliminary
results because using preliminary results can lead to misdiagnosis.
This symbol... Indicates that...
↑ the result is above the patient expected range.
Ø the result is below the patient expected range.
-----↑ the result is above the reporting range.
-----↓ the result is below the reporting range.
# the system detects the presence of substances in the sample that
may interfere with measurement.
-----? the system has an atypical response when measuring this parameter
and cannot report the result.
<
the result is less than the analytical range limit.
>
the result is greater than the analytical range limit.