5-18 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
d. Squeeze the finger snaps on the reagent manifold and then remove the manifold
by sliding it down and away from the sample connector.
Figure 5-8 Reagent Manifold Finger Snaps
e. Dispose of the manifold according to your institution’s protocol.
4. Install the new reagent manifold:
a. Snap the new reagent manifold in place.
NOTE: When you are installing the tubing, avoid pinching or bending the tubing.
b. Connect the waste tubing.
c. For Rapidlab 1245 and 1265 systems, connect the reagent manifold tubing to the
CO-ox pump tubing at the fitting.
Completing Reagent Manifold Maintenance
1. Close the front cover.
2. Reinstall the reagent and wash cartridges.
3. Select
Return.
4. Verify system performance:
a. Perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
b. Analyze a minimum of 2 levels of quality control material.
1 Finger snaps