Rapidlab 1200 Operator’s Guide: System Specifications E-5
02087462 Rev.V
The following table lists the units, reporting ranges, and display resolutions for the
parameters that can be entered into the Rapidlab
1200 systems.
Understanding System Limitations
The manufacturer cannot guarantee system performance when any of the
following situations occur. Specific terms of warranty, service, and contract
agreements may be invalidated if any of these situations occur.
you exceed the expiration dates of cartridges
you do not use cartridges according to the manufacturers recommendations
you do not follow standard laboratory practices
you do not follow the procedures described in this guide
you do not follow the environmental operating conditions and location
recommendations
Table E-8 Entered Parameters
Parameter Units Reporting Range Resolution
Temperature °C
°F
10.0–43.9
50.0111.0
0.1
0.1
F
I
O
2
%
decimal fraction
15.0–100.0
0.150–1.000
0.1
0.001
ctHb g/dL
g/L
mmol/L
2.0–25.0
20–250
1.2–15.5
0.1
1
0.1
Qt L/min 0.00–30.00 0.01
OBF 1.35–1.40 0.01
Flow L/min 0.00–99.99 0.01
Resp Rate b/min 0.0–100.0 0.1