EasyManua.ls Logo

Siemens Rapidlab 1200 - Page 364

Siemens Rapidlab 1200
448 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E-4 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
The following table lists the units, reporting ranges, and display resolutions for the
temperature corrected and respiratory parameters calculated by the Rapidlab
1200
systems.
The following table lists the units, reporting ranges, and display resolutions for the
metabolic parameters calculated by the Rapidlab
1200 systems.
ctO
2
(Hb) mL/dL
mL/L
mmol/L
0–40.0
0–400
0–17.9
0.1
1
0.1
Table E-6 Temperature Corrected and Respiratory Parameters
Parameter Units Reporting Range Resolution
pH(T) pH 6.000–8.000 0.001
H
+
(T)
nmol/L 10.0–1000.0 0.1
pCO
2
(T) mmHg
kPa
5.0–250.0
0.67–33.33
0.1
0.01
pO
2
(T) mmHg
kPa
0.0–800.0
0.00–106.66
0.1
0.01
pO
2
(A-a)(T) mmHg
kPa
0–999.9
0.00–133.31
0.1
0.01
pO
2
(a/A)(T) %
decimal fraction
0–100
0.00–1.00
1
0.01
RI(T) %
decimal fraction
0–2000
0.00–20.00
1
0.01
sp/ t(T)
%
decimal fraction
0–100
0.00–1.00
1
0.01
sp/ t(est, T)
%
decimal fraction
0–100
0.00–1.00
1
0.01
Table E-7 Metabolic Parameters
Parameter Units Reporting Range Resolution
HCO
3
-
act
mmol/L 0.0–99.9 0.1
HCO
3
-
std
mmol/L 0.0–99.9 0.1
ctCO
2
mmol/L 0.0–99.9 0.1
BE(B) mmol/L -99.9–99.9 0.1
BE(ecf) mmol/L -99.9–99.9 0.1
Table E-5 CO-ox Module Parameters
Parameter Units Reporting Range Resolution

Table of Contents

Related product manuals