Rapidlab 1200 Operator’s Guide: System Specifications E-3
02087462 Rev.V
The following table lists the units, reporting ranges, and display resolutions for
parameters measured by the Rapidlab
1200 systems CO-ox module.
ctO
2
([a- ]/a)
%
decimal fraction
0–100
0.00–1.00
1
0.01
O
2
mL/min
L/min
mmol/min
0–3500
0.00–3.50
0–156.2
1
0.01
0.1
O
2
mL/min
L/min
mmol/min
0–3500
0.00–3.50
0–156.2
1
0.01
0.1
Table E-5 CO-ox Module Parameters
Parameter Units Reporting Range Resolution
tHb g/dL 2.0–25.0 0.1
g/L 20–250 1
mmol/L 1.2–15.5 0.1
FO
2
Hb % (blood)
% (QC)
0.0 100.0
0.0 200.0
0.1
0.1
fraction (blood)
fraction (QC)
0.0001.000
0.000 ± 2.000
0.001
0.001
FCOHb % (blood)
% (QC)
0.0100.0
0.0 ± 200.0
0.1
0.1
fraction (blood)
fraction (QC)
0.0001.000
0.000 ± 2.000
0.001
0.001
FMetHb % (blood)
% (QC)
0.0100.0
0.0 ± 200.0
0.1
0.1
fraction (blood)
fraction (QC)
0.0001.000
0.000 ± 2.000
0.001
0.001
FHHb % (blood)
% (QC)
0.0100.0
0.0 ± 200.0
0.1
0.1
fraction (blood)
fraction (QC)
0.0001.000
0.000 ± 2.000
0.001
0.001
sO
2
% 15.0–100 0.1
fraction 0.15–1.000 0.001
Hct % 6–74 1
fraction 0.06–0.74 0.01
BO
2
(O
2
CAP) mL/dL 0.0–40.0 0.1
mL/L 0–400 1
mmol/L 0.0–17.9 0.1
Table E-4 Oxygenation Parameters
Parameter Units Reporting Range Resolution