8-42 Rapidlab 1200 Operator’s Guide: System Configuration
02087462 Rev. V
You need a formatted 1.44 MB 3.5-inch diskette or 1 CD to complete this activity.
1. At the Setup screen, select Secured Options.
2. Select the appropriate drive (disk or CD).
3. When prompted, insert a disk or CD into the appropriate drive and select
Continue.
4. When prompted, remove the diskette or CD from the drive and select
Continue.
5. Store the diskette or CD in a safe place, away from heat and strong magnetic sources
such as centrifuges.
Installing New System Software
You can install new system software from a CD-ROM, disks, or from a network
connection. To install software from a network connection, your system must be
configured for a network and be connected to a device that provides access to the new
system software.
You need the following materials to perform this task:
Disks containing the Rapidlab 1200 system software
CD-ROM containing the Rapidlab 1200 system software
You can copy patient, QC, and calibration data to a disk or CD. Refer to Copying Data
Files‚ page 7-2.
If you want to... select...
save Setup data,
Save Setup.
restore Setup data or copy Setup data from another system,
Restore Setup.