Rapidlab 1200 Operator’s Guide: Quality Control 4-11
02087462 Rev. V
To print the Recall–QC Statistics data, select the Print button. The printed report
contains the following additional information:
• parameter labels and statistics for all parameters
• measured and target values for each parameter
• the difference between the measured and target values, in percentage
• if a parameter exceeds the RiliBÄK range, text indicates the parameter has
failed
Using the Recall–Levey-Jennings Screens
The Recall–Levey-Jennings screens allow you to select a particular lot of QC
material and a parameter and then to view a month’s worth of data as a
Levey-Jennings graph.
NOTE: You must have defined range values for the system to generate a
Levey-Jennings graph.
1. At the user interface, select
Recall.
2. At the Recall screen, select QC, and then select
Levey-Jennings Graph.
3. At the Recall–Levey-Jennings Graph screen, select the lot number for the QC
material.
Use the up and down scroll arrows to navigate the list. The list is in
alphanumeric order, with the AutomaticQC lots listed first as AQC1, AQC2,
and so on.
4. Select a month.
The current month is uppermost of the three buttons. The 2 previous months
are next in order.
5. Select a parameter.
6. Select
Continue.
This displays the Recall Levey-Jennings Graph screen.
Viewing the Recall–Levey-Jennings Graph Screen
The Levey-Jennings graph displays a visual representation of the performance of
the selected parameter relative to QC measurements for up to a 31-day period of a
single month.
Select the Print button to print the Levey-Jennings graph. You can only print the
Levey-Jennings graph to an external printer.
If you want more than one month of data, select and print one month and then
select and print another month.
Under the banner, the Recall–Levey-Jennings Graph screen displays the lot
information for the currently displayed lot of QC material:
• parameter label for the data displayed in the graph