4-10 Rapidlab 1200 Operator’s Guide: Quality Control
02087462 Rev. V
The mean value is defined as the sum of all results divided by the number of results,
according to the following formula:
SD(R), standard deviation with RiliBÄK range checking defined according to the
following formula:
SD(R) as a percentage, standard deviation as a percentage with range checking
defined according to the following formula:
SD( ), standard deviation against the mean is defined according to the following
formula:
CV, coefficient of variation is defined according to the following formula:
N, number of samples included for these statistics.
NOTE: For RMSD and RMSD%, a minimum of 15 samples is required. If fewer than
15 samples are analyzed, the field is blank. For Standard Deviation and Coefficient of
Variation, a minimum of 5 samples is required. If fewer than 5 samples are analyzed,
the field is blank.
NOTE: If a sample is out of range it may skew QC Analysis results. To exclude a
sample you suspect is out of range, see Using the QC List Screen‚ page 4-7.
NOTE: Low and high AQC and RQC ranges are adjusted to adhere to the
RiliBÄK 2008 requirements that define deviation from target ranges.
NOTE: CO-ox fractions for FO
2
Hb, FCOHb, FMetHb, and FHHB are not adjusted.
N
xxx
x
N
x
N
i
N
i
+
+
+
=Σ=
=
L
21
1
1
2
arg
1
)(
1
ett
n
i
i
xx
n
RMSD =
=
100*)/(%
arg ett
xRMSDRMSD
=
x
2
1
)(
)1(
1
xx
N
S
i
N
i
Σ
=
=
%100×=
x
S
CV