6-46 Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
Performing the Wavelength Cal Test
The Wavelength Cal test verifies the proper operation of the CO-ox neon lamp and prints
the results providing the system ID, the date and time of the report, and as represented in
Table 6-7.
To perform the Wavelength Cal Report, select CO-ox > Wavelength Cal.
Performing the CO-ox Sample Chamber Test
This test is only performed on Rapidlab 1245 and 1265 Systems.
Use this procedure to test the ability of the CO-ox sample chamber to move to the open
and close positions. The systems displays the pass or fail results on a printed report.
1. Select Status > Diagnostics.
2. Enter your password, if necessary.
3. Select
CO-ox > Start.
The sample chamber moves to the open position.
4. Select
Change to move the sample chamber to the closed position.
5. Select
Stop.
The system prints the test results Passed or Failed.
6. Wait while the system returns the sample chamber to the correct position.
7. Select
Return.
8. To return to the Status screen, select
Return.
Table 6-7 Wavelength Cal Report
Row number Data
1 Mean
2 Standard deviation
3 Measured position
4 Baseline position
5Shift
6 Integration time from last zero
7 Serial number