5-52 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Flushing the Upper Connector
1. Connect the syringe tubing to the upper connector for the wash reagent.
Figure 5-26 Reagent Manifold
2. Flush with a minimum of 3 mL water.
3. Flush with a minimum of 12 mL air.
4. Collect the water in an absorbent swab or equivalent.
1 Water collection area
2 Upper connector