Rapidlab 1200 Operator’s Guide: Quality Control 4-7
02087462 Rev. V
Recalling QC Results
Use the Recall screen to view and print results for previously analyzed QC
samples. When you select the
Recall button and then select QC, the Recall QC
screen displays the following options:
•QC List
• QC Statistics
• Levey-Jennings Graph
Using the QC List Screen
The QC List screen displays the QC results that are stored in the system. The list
box on the QC List screen displays Required QC, Unscheduled QC,
AutomaticQC, and High G/L results.
1. At Analysis screen, select Recall, and then select QC.
2. At the Recall QC Options screen, select
QC List.
This displays the Recall QC Data screen.
• For AutomaticQC samples, the list uses AQC for the control name and
uses the last 6 digits of the cartridge serial number as the lot number.
• For Required QC samples, the system displays an arrow symbol next to
the date.
3. Locate the sample results you want to view:
NOTE: You can use the Search to find a specific result or to filter the results
according to criteria you select. Refer to Using the Recall QC Data Search
Screen‚ page 4-8.
a. Use the up or down arrow buttons to scroll through the screen of results.
b. Use the horizontal scroll arrows to view the Parameter values for the
selected sample result.
4. Select the sample and then select
Results.
5. At the Results screen,
you can perform the following tasks:
6. Select
Continue.
To do this... Select this...
reject results, You cannot undo this function. After you reject the
results, you cannot accept them again.
Reject Results.
The results are excluded from QC Statistics
calculations and plotted differently on
Levey-Jennings graphs.
print result, the
Print button.
send results to a data
management system or LIS,
Send.