Rapidlab 1200 Operator’s Guide: Maintenance 5-9
02087462 Rev. V
Performing Every 60 Days Maintenance
The every 60 days maintenance schedule is based on analyzing 30 samples per
day, unless otherwise noted. If your laboratory analyzes more than 30 samples per
day, perform this maintenance more frequently.
Replacing the CO-ox Sample Chamber
You need a new CO-ox sample chamber to perform this task.
BIOHAZARD: Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for
recommended
precautions when working with biohazardous materials.
1. Select
Status > Maintenance > Replacing the CO-ox Sample Chamber.
2. To start the procedure, select
Continue.
3. Follow the instructions in the video.
4. Allow the system to warm up for 5 minutes.
5. Perform a full calibration.
Refer to Recalling Calibration Results‚ page 3-4.
6. Analyze a minimum of 2 levels of quality control material.
Checking the Air Filters
1. Select Status > Maintenance > Replacing the Air Filter.
2. Select
Continue.
3. To remove the filters, follow the instructions in the video.
4. Shake the filters to remove dust.
If the filters are torn or cannot be cleaned, replace the filters.
5. To reinstall the filters, follow the instructions in the video.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Rapidlab 1200 and is the answer not in the manual?

Siemens Rapidlab 1200 Specifications

General IconGeneral
BrandSiemens
ModelRapidlab 1200
CategoryMedical Equipment
LanguageEnglish

Related product manuals