Rapidlab 1200 Operator’s Guide: Troubleshooting 6-33
02087462 Rev. V
b. Push the clot-removal line through the fluid connector on the reagent
manifold.
Figure 6-17 CO-ox Sample Path to Waste
c. Re-connect the CO-ox pump tubing.
5. Reinstall the cartridges and close both the reagent and wash doors.
6. Perform the RCx reagent flow test:
a. Select
Back > Fluidic Functions > RCx.
b. Look at the sample path for obstructions and flow problems.
7. Return to the Status screen to view the events log.
8. Verify sensor performance:
a. Perform a 2-point calibration.
b. Analyze a minimum of 2 levels of quality control material.
Using Diagnostics
NOTE: Operators with level 1 or level 2 security access can access the diagnostic
tests.
NOTE: The system cannot perform routine fluidic activities while the Diagnostics
screen or subscreens are displayed. When you exit from the Diagnostics screen,
the system may perform an extended calibration before you can use the system.
1 Pump tubing at the fitting.
2 Fluid connector.