6-32 Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
b. Rotate the gray wheel towards you, pull the tubing through the fluid detector, and
remove the tubing from the steel measurement module connector.
c. Push the new tubing through the fluid detector.
d. Connect the tubing to the steel measurement module connector and to the sample
chamber connector.
Use a clamp or tweezers to help you attach the tubing.
3. Remove obstructions from the CO-ox waste tubing:
a. Disconnect the tubing from the steel fluid connector and the connector in the
pump tubing.
Figure 6-16 CO-ox Sample Path
b. Push a clot-removal line through the tubing, opposite the direction of sample flow.
c. Reconnect the CO-ox waste tubing.
4. Remove obstructions from the waste path in the reagent manifold:
a. Disconnect the CO-ox pump tubing from the reagent manifold tubing at the
fitting.
1 Steel fluid connector