Rapidlab 1200 Operator’s Guide: Quality Control 4-9
02087462 Rev. V
Using the Recall–QC Statistics Screens
The Recall–QC Statistics screens allow you to select a particular lot of QC
material and then to view all the statistics for that lot.
To access the Recall–QC Statistics screens, perform the following steps:
1. At the Analysis or Status screens, select Recall.
2. At the Recall screen, select
QC and then select QC Statistics.
3. At the Recall–QC Statistics screen, select the AQC Level or lot ID number for
the QC material.
Use the up and down scroll arrows to navigate the list. The list is in
alphanumeric order, with the AutomaticQC lots listed first as AQC1, AQC2,
and so on.
To narrow the date range for samples being analyzed, enter dates in the Start
and End date boxes, using the numeric keypad.
4. Select
Continue.
A Wait screen displays during data analysis. When data analysis completes,
the QC Statistics screen displays QC Statistics.
The Recall–QC Statistics screen only displays accepted results. Any results
that are reported out of the range of the instrument or questionable are
excluded from these statistics.
Viewing the Recall–QC Statistics Screen
Under the banner, the Recall–QC Statistics screen displays the following lot
information for the currently displayed lot of QC material:
QC material label, RapidQC Complete or AutomaticQC
QC lot level
QC lot number (does not display for Automatic QC)
start date, the earliest date this lot was used according to these statistics
end date, the most recent date this lot was used according to these statistics
In the QC Statistics list, the screen displays the following statistics for each
measured parameter:
Mean Value