Rapidlab 1200 Operator’s Guide: Maintenance 5-19
02087462 Rev. V
Replacing the AutomaticQC Manifold
You need an AutomaticQC manifold for this task.
1. Press down on the left side of the AutomaticQC cartridge connector and slide
the connector to the right.
2. Select
Continue.
3. Select
Diagnostics > Cartridges.
4. Select
Eject R Cartridge, remove the reagent cartridge, and set it aside.
5. Lift the front cover.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
6. Remove the AutomaticQC manifold:
a. Pull the connector port cover towards you.
Figure 5-9 Connector Port and Cover
b. Push down on the pins on the manifold and pull the manifold and
connector port away.
7. Install the new AutomaticQC manifold:
a. Insert the bottom clip into the opening on the system.
1 Connector port cover
2 Cartridge connector
3 Pins