Rapidlab 1200 Operator’s Guide: Maintenance 5-45
02087462 Rev. V
10. Start up the system:
a. Turn the power switch on.
b. At the interface screen, use the numeric buttons to enter your password.
c. Select
Continue.
Replacing the CO-ox Roller Cage
Use this procedure to replace the roller cage of the CO-ox pump.
You need the following materials for this task:
• Lint-free tissue and swabs
• Roller cage
•Water
Removing the Roller Cage
1. Select Status > Maintenance.
2. Lift the front cover.
Refer to Preparing for Maintenance Procedures‚ page 5-1.
3. Disconnect the reagent manifold tubing from the CO-ox pump tubing at the
fitting.
Figure 5-22 CO-ox Module and Reagent Manifold
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
1 Fitting
2 Reagent manifold tubing