5-16 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Replacing the CO-ox Pump Tubing
1. Disconnect the CO-ox pump tubing from the CO-ox waste tubing.
Figure 5-6 CO-ox Module and Reagent Manifold
2. Disconnect the CO-ox pump tubing from the reagent manifold tubing at the fitting.
3. Do not disconnect the reagent manifold tubing from the reagent manifold.
4. Grasp the lower tubing cuff and pull the tubing away from the platen and the roller
cage.
5. Place the lower tubing cuff of the new tubing under the lower side of the platen, press
the tubing around the outside of the rollers, and turn the roller cage clockwise until the
tubing rests on the rollers.
6. Connect the tubing to the waste tubing and the reagent manifold tubing.
Completing CO-ox Maintenance
1. Close the front cover.
2. Reinstall the wash cartridge.
3. Dispose of the tubing according to your institution’s protocol.
4. Select
Return.
5. Continue performing yearly maintenance or perform a full calibration.
1 Roller cage
2 Platen
3 Lower tubing cuff
4 Reagent manifold tubing