6-62 Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
D23 Reagent Flow Error: 6
D23 Reagent Flow Error: 7, 8
Problem Corrective Action
The system detects
an air leak in the
fluidic path during
a calibration.
1. Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
2. Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3. If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
4. If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
5. If the problem persists, call for technical assistance.
Problem Corrective Action
Wash fluid was not
detected during
wash after a
calibration or
sample analysis.
7 calibration
8 sample analysis.
1. Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
2. Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3. If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
4. If the problem persists, call for technical assistance.