8-28 Rapidlab 1200 Operator’s Guide: System Configuration
02087462 Rev. V
c. Select Printer and Devices > Communications.
d. Select
RS-232 > Configure.
e. Select the following options for the communication settings:
f. Select
Save.
6. At the Communications screen, select
Save.
7. When prompted that the connection was unsuccessful, select
Continue.
8. At the Rapidlink or Rapidcomm server or workstation, enter the communication
settings given above.
Refer to the Rapidlink Supervisors Manual or Rapidcomm Users Guide.
9. At the Rapidlab 1200 system, select
Save.
10. Select
Return.
11. To complete any additional setup that may be necessary to operate the Rapidlab 1200
system with the Rapidlink or Rapidcomm program, refer to the Rapidlink Supervisor’s
Manual or Rapidcomm Users Guide.
Connecting to a CompleNet
®
Network Connection
CompleNet network connection is a proprietary protocol that allows real-time
communication between a Rapidlab
1200 system and a Rapidlink or Rapidcomm
workstation. Use this protocol to connect your system to a Rapidlink or Rapidcomm
workstation if your Rapidlink or Rapidcomm system is part of an Ethernet network that
uses a 10Base-T connection.
You need the following materials for this activity:
Twisted-pair cable with RJ-45 connector
Rapidlink Supervisors Manual or Rapidcomm Users Guide
Select 1 of the following methods to configure the network connection:
Enter IP addresses
Use DHCP (dynamic host configuration protocol)
Setting Default Selection
Baud 9600
Parity Even
Data Bits 8