Rapidlab 1200 Operator’s Guide: System Configuration 8-5
02087462 Rev. V
6. At the Required QC Ranges screen, select Save to return to the Setup menu.
Editing Required QC Ranges
You can change the range values of each parameter in the selected level. You can
also convert a trial lot to an active lot.
1. At the Setup screen, select QC > Required QC Ranges.
2. Select an available level button to view or edit the ranges for that level:
Selecting the level button displays the Range Entry screen displaying the
ranges defined for each parameter for that level.
3. Select the parameter you want to edit and then select
Low or High.
4. To clear the field, select
Clear.
5. Use the numeric buttons to enter new range values.
6. Repeat steps 3 through 5 for each parameter you need to edit.
7. If the parameter is a trial lot and it should be an active lot, select
Set Active.
The Set Active button is only displayed when the lot is a trial lot.
8. Use the print button to print the values selected for this level.
9. Select
Save.
Setting up AutomaticQC Schedule
Use this screen to select the days and times for AutomaticQC analysis and select
the controls to analyze at the scheduled times.
1. At the Setup screen, select QC > AutomaticQC Schedule.
If a schedule is already defined, the AutomaticQC Schedule screen displays
the currently defined schedule.
2. Select each button for the level you want to schedule.
a. Use the arrow keys to navigate to a place in the grid indicating the day of
the week and time period for the level or levels.
b. Select one or more buttons for the level or levels to schedule.
A check mark on the button means that you scheduled that level. Select
the button again to remove the check mark.
3. Turn AutomaticQC analysis on:
a. Select
QC > QC Options > AutomaticQC.
b. Select
Repeat if you want the system to repeat QC analysis when a
parameter fails the first analysis.
c. Select
Save.