Rapidlab 1200 Operator’s Guide: Troubleshooting 6-53
02087462 Rev. V
D4 Codes
D4 Offset Error: pO
2
pCO
2
D4 Offset Error: pH Na
+
K
+
Ca
++
Cl¯
Problem Corrective Action
Sensor offset is
beyond predefined
limits during a
1-point or a 2-point
calibration.
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
Perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
Check that the sensors are installed in the correct location
and are aligned, the O-rings are in place, and the
spring-loaded latch is closed.
If necessary, remove and install the sensors correctly.
If the problem persists, replace the sensor.
If the problem persists, call for technical assistance.
Problem Corrective Action
Sensor offset is
beyond predefined
limits during a
1-point or a 2-point
calibration.
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
Perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
Check the solution level in the affected measurement sensor
and look for bubbles.
The K
+
, Ca
++
, and Cl¯ sensors should be nearly full. The pH
and Na
+
sensors should be full. If the solution level is low,
refill the sensor.
•If K
+
or Ca
++
have an offset error, replace the fill solution
even if the fluid levels are sufficient.
If more than one sensor has the offset error, check the
solution level in the reference sensor and look for salt in the
vent hole and bubbles in the sensor.
Clean and refill the reference sensor as needed.
Check that the sensors are installed in the correct location
and are aligned, the O-rings are in place, and the
spring-loaded latch is closed.
If necessary, remove and install the sensors correctly.
If the problem persists, replace the sensor.
If the problem persists, call for technical assistance.