Rapidlab 1200 Operator’s Guide: Troubleshooting 6-61
02087462 Rev. V
D23 Reagent Flow Error: 4, 12
D23 Reagent Flow Error: 5, 13
Problem Corrective Action
Slope Cal reagent
was not detected
during calibration.
4 FD1
12 FD2
1. Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
2. Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3. If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
4. If the problem persists, call for technical assistance.
Problem Corrective Action
RCx reagent was
not detected during
calibration.
5 FD1
13 FD2
1. Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
2. Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3. If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
4. If the problem persists, call for technical assistance.