Rapidlab 1200 Operator’s Guide: System Configuration 8-3
02087462 Rev. V
Entering Customized QC Range Limits
1. At the Setup screen, select QC.
2. If performing AQC, select
AutomaticQC Ranges.
If performing RQC, select
Required QC Ranges.
3. Select the QC range to view.
The screen for the QC range you selected displays.
4. Select a parameter.
5. Enter a low limit, a high limit, or both, in the Low and High boxes, for the
selected parameter.
If in RiliBÄK mode, use the following equations to determine low and high
values:
• Range Offset from Manufacturing Target =
Manufacturing Target + (Manufacturing Target * Desired Deviation from
Target)
•QC Low Value =
(Manufacturing Target – Range Offset from Manufacturing Target)
•QC High Value =
(Manufacturing Target + Range Offset from Manufacturing Target)
6. Repeat steps 4-5 as needed, for additional parameters.
7. Select
Continue.
Setting Up Required QC
Use the Required QC screen to set up a required QC sample schedule. Use the
barcode scanner and RapidQC controls to perform Required QC analysis.
NOTE: Ensure that you enter information for each lot of the controls you
schedule. Refer to Setting up Required QC Ranges‚ page 8-4. You cannot analyze
the scheduled controls until you enter the lot information.
1. At the Setup screen, select
QC > Required QC Schedule.
The screen displays the Required QC Schedule with the current schedule for
Required QC analysis, if defined.
NOTE: When you change intervals, the system clears any times and levels
that were previously scheduled.
2. Select
Intervals.
The system displays the Intervals screen.
3. Select the number of hours for the interval of the Required QC analysis.
For example, select 8 to have an 8-hour window for performing scheduled
controls.