2-6 Rapidlab 1200 Operator’s Guide: Operating the System
02087462 Rev. V
Shutting Down the System
Use this procedure to remove power from the system for 24 hours or less.
CAUTION: Do not remove power from the system for more than 6 hours if
cartridges are installed. Cartridges installed in the system remain stable for 6
hours without power.
CAUTION: To prevent damage to the hard drive and permanent loss of data, use
this procedure to shut down the system.
1. Select
Status > Shutdown > Yes.
2. When prompted, turn the power switch off.
The power switch is on the back panel of the system.
3. To restore power to the system, turn the power switch on.
WARNING: Do not operate the system in the presence of flammable anesthetic
mixture with air, O
2
, or nitrous oxide. The risk of explosion exists in a potentially
explosive environment. Refer to Protecting Yourself from Electrical Hazards‚
page A-3 for recommended precautions when working around electricity.