Rapidlab 1200 Operator’s Guide: Troubleshooting 6-17
02087462 Rev. V
16. Select Yes to calibrate the system.
The clot clearing procedure is complete.
Clearing the Clot from the Preheater
Use this procedure only if the Clearing the Clot from the Reagent Manifold
procedure fails to clear the clot. The Clearing the Clot from the Preheater
procedure continues from step 11 on page 16.
1. Open the measurement module cover, by lifting the 2 latch buttons located at
the bottom left and right of the module. See Figure 6-3.
Figure 6-3 Measurement Module Compartment Cover
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to Appendix A, Protecting Yourself from Biohazards for recommended
precautions when working with biohazardous materials.
1 Measurement module latch buttons
CAUTION: Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.