6-2 Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
Parameter
failed
AutomaticQC
analysis
because the
sensor or the
cartridge is not
working
correctly.
Perform the following steps until the system is
ready for use:
Perform up to two 2-point calibrations.
If the parameter fails the 2-point calibrations,
view the events log for D codes and messages.
Replace the sensor if necessary.
If the parameter passed the 2-point
calibrations, perform AutomaticQC analysis
for the failed level of control.
If the parameter fails AutomaticQC analysis,
analyze a QC sample for the failed level using
QC material from an ampule.
If the parameter fails the ampule QC analysis,
review the events log for D codes and
messages.
If the parameter passes the ampule QC
analysis, replace the AutomaticQC cartridge.
The target
ranges for one
or more
parameters are
too narrow.
Verify that the target ranges entered for the
control are correct.
If you specify target ranges that are narrower
than those provided with the control,
parameters may fail QC analysis.
Analyze QC for the failed level.
Refer to Performing Unscheduled QC Sample
Analysis‚ page 4-4.
If the results are within range, restore the
parameter.
Refer to Restoring Parameters‚ page 4-12.
For AutomaticQC samples, analyze the failed
level.
Refer to Performing Unscheduled QC Sample
Analysis‚ page 4-4.
If the results are within range, the system
restores the parameter.
If analysis fails again, review the events log for
D codes and messages.
Error
Probable
Cause Corrective Action