Rapidlab 1200 Operator’s Guide: Operating the System 2-33
02087462 Rev. V
Sample Totals Prints a report of the total number of
patient and QC samples performed on the
system.
Refer to Viewing the Sample Totals‚
page 7-4.
Last Cal Reports Prints data for each sensor from the last
2-point or full calibration.
Refer to Using the Sensors Screen‚
page 6-44.
Printer Report Prints all of the printable characters in the
printer memory.
Refer to Performing Tests and Printing
Diagnostic Reports‚ page 6-34.
Temperature/pAtm Report Prints an index and one snapshot of the
heater readings.
Refer to Performing Tests and Printing
Diagnostic Reports‚ page 6-34.
Lamp Calibration Report Prints a Pass/Fail value for the CO-ox
lamp calibration. Only available on
Rapidlab 1245 and 1265 systems.
Refer to Performing the Lamp Calibration
Test ‚ page 6-45.
Wavelength Calibration Report Prints the result of the Wavelength
Calibration test. Only available on
Rapidlab 1245 or 1265 systems.
Performing the Wavelength Cal Test‚
page 6-46.
Sample Chamber Report Prints a Pass/Fail value for the result of
the Sample Chamber test. Only available
on Rapidlab 1245 and 1265 systems.
Refer to Performing the CO-ox Sample
Chamber Test‚ page 6-46
Voltages and Temperature Report Prints the historical Pass/Fail condition of
all voltages and temperatures.
Refer to Performing Tests and Printing
Diagnostic Reports‚ page 6-34.
Leak Test Report Prints the result of the Leak Test.
Refer to Performing the Leak Test‚
page 6-37.
Measurement Test Report Prints measurement readings for all
sensors.
Refer to Performing Tests and Printing
Diagnostic Reports‚ page 6-34.
Report Name Description