5-38 Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
5. Fill the KCl reservoir with KCl fill solution:
a. Remove the reservoir cap using the hex tool and set the cap aside.
b. Partially fill the KCl reservoir.
c. Tap the front face of the sensor with your knuckle to remove any bubbles.
d. Continue filling the reservoir with KCl fill solution to the fill line.
6. Hand tighten the reservoir cap.
7. Clean and inspect the sensor:
a. Tap the front face of the sensor with your knuckle to remove any bubbles.
b. Wipe any excess fill solution from the exterior of the sensor using a lint-free
tissue.
c. Ensure that the O-rings are in place.
8. Install the Reference sensor.
Refer to Installing the Sensor‚ page 5-31.
9. Verify sensor performance.
Refer to Verifying Sensor Performance‚ page 5-31.
CAUTION: Do not overtighten the reservoir cap. Overtightening can deform the
gasket and cause leaks.