Rapidlab 1200 Operator’s Guide: Troubleshooting 6-21
02087462 Rev. V
b. Reinsert the reagent manifold tubing (item 2, Figure 6-7) into the fitting
(item 1, Figure 6-7) that connects to the CO-ox pump tubing
(Rapidlab 1245 and 1265 systems).
c. Reinsert the waste tubing into the port on the top, right-side of the reagent
manifold (item 4, Figure 6-7).
If an obstruction has been cleared, go to step 9.
If an obstruction has not been cleared, go to Clearing the Clot from
the Reagent Cartridge‚ page 6-22
.
9. Reinstall the pO
2
sensor.
a. Push the black spring enclosure assembly to the right, until the
mechanism clicks in place.
b. Remove the syringe.
c. Remove the clot clearing tool, by lifting the clot clearing tool lip up at the
bottom while pressing the top of the tool inward.
d. Reinstall the pO
2
sensor.
e. Press the black spring enclosure assembly lever to reapply pressure to the
sensor assembly.
10. Close the system.
a. Reinstall the reagent and wash cartridges.
b. Close the reagent and wash module doors.
c. Close the front cover of the system.
11. Select the Return button twice to exit Diagnostics mode.
You are prompted to calibrate the system.
12. Select
Yes to calibrate the system.
The clot clearing procedure is complete.