Rapidlab 1200 Operator’s Guide: Troubleshooting 6-25
02087462 Rev. V
NOTE: Use gauze to catch waste material cleared by the clot-removal
line.
2. To remove obstructions from the sample tubing in the reagent manifold, push
the clot-removal line through the fluid connector on the reagent manifold.
Figure 6-9 Sample Tubing in the Reagent Manifold
3. Remove obstructions from the preheater:
a. Remove the pO
2
and pCO
2
(gas) sensors.
1 Sample connector
2 Fluid connector