8-36 Rapidlab 1200 Operator’s Guide: System Configuration
02087462 Rev. V
When access is Restricted, the system requires a password for every level of operator for
all functions.
When access is Limited, some functions do not require a password.
Table 8-1 Restricted Access Setting
Level/Pass Level/Pass Level/Pass Level/Pass
Function 1234
analyzing samples Yes/Yes Yes/Yes Yes/Yes Yes/Yes
recalling data Yes/Yes Yes/Yes Yes/Yes Yes/Yes
replacing cartridges Yes/Yes Yes/Yes Yes/Yes No
initiating calibrations Yes/Yes Yes/Yes Yes/Yes No
accessing Setup
options
Yes/Yes Yes/Yes No No
accessing Diagnostics Yes/Yes Yes/Yes
No
No
accessing the Restore
QC screen
Yes/Yes Yes/Yes Yes/Yes No
accessing secured
options
Yes/Yes No No No
Table 8-2 Limited Access Setting
Level/Pass Level/Pass Level/Pass Level/Pass
Function 1234
analyzing samples Yes/No Yes/No Yes/No Yes/No
recalling data Yes/No Yes/No Yes/No Yes/No
replacing cartridges Yes/Yes Yes/Yes Yes/Yes No
initiating calibrations Yes/Yes Yes/Yes Yes/Yes No
accessing Setup
options
Yes/Yes Yes/Yes No No
accessing Diagnostics Yes/Yes Yes/Yes No No
accessing the Restore
QC screen
Yes/Yes Yes/Yes Yes/Yes No
accessing secured
options
Yes/Yes No No No