Rapidlab 1200 Operator’s Guide: System Specifications E-19
02087462 Rev.V
Precision on manual aqueous quality control materials was estimated using a
minimum of 3 Rapidlab 1245 systems. Controls were measured over a minimum
of 10 days.
Precision on the aqueous quality control materials in the AutomaticQC module
was estimated using a minimum of 3 Rapidlab 1245 systems over a minimum of
10 days. AutomaticQC is automatically scheduled to run (single sample)
independently 3 times per day.
The following tables summarize the results of the Rapidlab 1245 system precision
for manual QC and AutomaticQC:
Table E-16 Rapidlab 1245 Manual QC Precision Results
Parameter Level n Mean WRSD
*
* WRSD = within-run standard deviation
Total SD
†
pH 1 161 7.119 0.001 0.002
pH 2 161 7.328 0.001 0.002
pH 3 160 7.529 0.001 0.003
pCO
2
1 161 72.6 0.42 1.14
pCO
2
2 162 40.7 0.29 0.56
pCO
2
3 160 22.9 0.21 0.36
pO
2
1 161 147.3 0.69 1.85
pO
2
2 162 97.7 0.85 1.56
pO
2
3 160 24.8 0.73 1.07
tHb 1 132 17.8 0.05 0.14
tHb 2 132 14.2 0.05 0.11
tHb 3 132 7.8 0.03 0.06
FO
2
Hb 1 132 80.0 0.15 0.21
FO
2
Hb 2 132 93.3 0.14 0.29
FO
2
Hb 3 132 60.1 0.22 0.32
FCOHb 1 132 1.0 0.14 0.19
FCOHb 2 132 1.8 0.11 0.22
FCOHb 3 132 16.4 0.18 0.26
FMetHb 1 132 15.9 0.11 0.15
FMetHb 2 132 2.0 0.10 0.21
FMetHb 3 132 7.1 0.18 0.41
FHHb 1 132 3.2 0.05 0.08
FHHb 2 132 2.9 0.05 0.11
FHHb 3 132 16.4 0.08 0.14