Rapidlab 1200 Operator’s Guide: System Specifications E-29
02087462 Rev.V
Micro-
syringe
168 0.941 5.553 7.223 0.990 26.2 259.5
Table E-27 Rapidlab 1245 System Human Whole Blood Method
Comparison—tHb
Mode N Slope Intercept RMSE r
2
tHb 296 0.963 0.542 0.513 0.991
Capillary 330 0.957 0.849 0.478 0.993
Syringe 296 0.960 0.671 0.471 0.993
Table E-28 Rapidlab 1245 System Human Whole Blood Method
Comparison—FO
2
Hb
Mode N Slope Intercept RMSE r
2
tHb 274 0.998 -0.123 2.292 0.994
Capillary 289 1.005 -0.892 1.701 0.997
Syringe 277 0.991 0.657 1.906 0.996
Table E-29 Rapidlab 1245 System Human Whole Blood Method
Comparison—FCOHb
Mode N Slope Intercept RMSE r
2
tHb 188 0.979 0.013 1.119 0.998
Capillary 209 0.996 -0.128 0.959 0.999
Syringe 211 0.988 0.174 0.972 0.999
Table E-30 Rapidlab 1245 System Human Whole Blood Method
Comparison—FMetHb
Mode N Slope Intercept RMSE r
2
tHb 202 0.990 0.055 1.055 0.999
Capillary 206 1.017 0.077 0.991 0.999
Syringe 210 0.970 0.054 0.951 0.999
Table E-31 Rapidlab 1245 System Human Whole Blood Method
Comparison—FHHb
Mode N Slope Intercept RMSE r
2
tHb 139 1.005 0.676 1.797 0.998
Capillary 146 0.994 0.808 1.634 0.998
Syringe 117 0.989 0.348 1.657 0.998
Table E-26 Rapidlab 1245 System Human Whole Blood Method Comparison—pO
2
Mode N Slope Intercept RMSE r
2
Min Max