Rapidlab 1200 Operator’s Guide: System Specifications E-31
02087462 Rev.V
Precision on manual aqueous quality control materials was estimated using a
minimum of 3 Rapidlab
1260 systems. Controls were measured over a minimum
of 10 days.
Precision on the aqueous quality control materials in the AutomaticQC module
was estimated using a minimum of 3 Rapidlab
1260 systems over a minimum of
10 days. AutomaticQC is automatically scheduled to run (single sample)
independently 3 times per day.
The following tables summarize the results of the Rapidlab 1260 system precision
for manual QC and AutomaticQC.
Table E-32 Rapidlab 1260 Manual QC Precision Results
Parameter Level n Mean WRSD
*
Total SD
Ca
++
1 154 1.68 0.007 0.011
Ca
++
2 155 1.31 0.004 0.008
Ca
++
3 152 0.83 0.006 0.008
Cl
-
1 152 78 0.42 0.67
Cl
-
2 153 98 0.34 0.74
Cl
-
3 150 117 0.44 1.15
Glu 1 156 195 1.33 3.01
Glu 2 157 95 0.92 1.51
Glu 3 154 50 0.54 1.02
K
+
1 156 3.24 0.005 0.013
K
+
2 157 5.34 0.011 0.027
K
+
3 154 7.24 0.012 0.042
Lac 1 114 11.1 0.13 0.97
Lac 2 116 1.0 0.05 0.05
Lac 3 114 3.0 0.05 0.22
Na
+
1 156 119.2 0.15 0.31
Na
+
2 157 143.8 0.21 0.42
Na
+
3 154 168.7 0.24 0.49
pCO
2
1 156 71.2 0.29 0.91
pCO
2
2 157 40.3 0.32 0.45
pCO
2
3 154 22.9 0.17 0.33
pH 1 155 7.118 0.001 0.003
pH 2 155 7.327 0.001 0.003
pH 3 153 7.527 0.001 0.003