Rapidlab 1200 Operator’s Guide: System Specifications E-47
02087462 Rev.V
Table E-44 Rapidlab 1260 Human Whole Blood Method Comparison—pCO
2
Mode N Slope Intercept RMSE r
2
Min Max
Capillary 215 0.960 2.066 3.430 0.997 6.9 226.2
Syringe 422 0.922 3.229 5.437 0.992 6.9 229.0
Micro-
capillary
108 1.056 -2.100 2.851 0.998 6.5 238.7
Micro-
syringe
215 1.017 -1.734 2.905 0.998 5.8 242.4
Table E-45 Rapidlab 1260 Human Whole Blood Method Comparison—pO
2
Mode N Slope Intercept RMSE r
2
Min Max
Capillary 197 0.996 2.965 6.092 0.999 28.5 681.6
Syringe 404 0.982 2.718 10.291 0.997 26.0 723.9
Micro-
capillary
108 0.987 1.114 6.417 0.999 24.9 684.7
Micro-
syringe
215 1.013 -3.218 7.534 0.999 24.9 714.8
Table E-46 Rapidlab 1260 Human Whole Blood Method Comparison—Na
+
Mode N Slope Intercept RMSE r
2
Min Max
Capillary 144 0.964 5.366 1.596 0.994 107.8 186.1
Syringe 287 0.987 2.153 1.352 0.996 107.6 184.3
Micro-
capillary
72 0.942 8.008 1.366 0.995 110.5 178.7
Micro-
syringe
141 0.979 3.552 0.806 0.998 109.8 182.6
Table E-47 Rapidlab 1260 Human Whole Blood Method Comparison—K
+
Mode N Slope Intercept RMSE r
2
Min Max
Capillary 137 1.066 -0.139 0.202 0.997 0.81 12.90
Syringe 270 1.095 -0.272 0.215 0.997 0.72 13.08
Micro-
capillary
72 0.962 0.165 0.132 0.999 1.26 12.35
Micro-
syringe
136 0.971 0.122 0.347 0.990 0.93 13.19