E-56 Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
1Micro-
capillary
12 0.22 6.8 7.1 95.2 3.2
2Micro-
capillary
12 0.28 21.3 21.3 99.7 1.3
3Micro-
capillary
12 0.24 28.1 28.4 99.0 0.9
4Micro-
capillary
12 0.18 35.9 35.8 100.3 0.5
5Micro-
capillary
11 0.36 48.9 49.8 98.3 0.7
6Micro-
capillary
12 0.53 70.2 71.1 98.9 0.8
7Micro-
capillary
12 1.69 96.4 100.2 96.2 1.7
8Micro-
capillary
12 3.48 133.2 135.1 98.6 2.6
9Micro-
capillary
12 3.73 221.7 229.1 96.8 1.7
1Micro-
syringe
24 0.21 6.8 7.1 96.8 3.1
2Micro-
syringe
24 0.64 21.5 21.4 100.4 3.0
3Micro-
syringe
24 0.45 28.3 28.6 99.1 1.6
4Micro-
syringe
24 1.11 34.3 35.7 96.2 3.2
5Micro-
syringe
24 0.96 47.6 49.5 96.3 2.0
6Micro-
syringe
24 1.84 71.1 71.4 99.5 2.6
7Micro-
syringe
24 4.35 95.6 99.8 95.7 4.6
8Micro-
syringe
24 1.42 130.9 134.3 97.4 1.1
9Micro-
syringe
24 5.21 217.7 228.2 95.4 2.4
* WRSD = within-run standard deviation
Table E-55 Rapidlab 1265 Recovery and Precision Testing—pCO
2
Level Mode N WRSD
*
Observed Expected Recovery WRCV