Rapidlab 1200 Operator’s Guide: System Specifications E-67
02087462 Rev.V
tHb Syringe 63 14.06 0.27 0.08
FO
2
Hb Capillary 62 85.68 -0.55 0.17
FO
2
Hb tHb 63 86.21 -0.54 0.30
FO
2
Hb Syringe 63 85.57 -0.30 0.26
FCOHb Capillary 62 5.29 0.05 0.15
FCOHb tHb 63 4.98 0.28 0.22
FCOHb Syringe 63 4.69 0.34 0.21
FMetHb Capillary 62 5.08 -0.05 0.05
FMetHb tHb 63 4.86 -0.23 0.08
FMetHb Syringe 63 5.39 -0.37 0.07
FHHb tHb 63 3.94 -0.05 0.22
FHHb Capillary 62 3.96 0.07 0.11
FHHb Syringe 63 4.35 -0.38 0.24
* tHb mode performance is equivalent to sample device used
† WRSD: within-run standard deviation
Table E-65 Rapidlab 1265 CO-ox Module Recovery and Precision with Total
Hemoglobin of ~20 g/dL and Oxyhemoglobin > 75%
Analyte Mode
*
n
Mean
Reference Mean Bias WRSD
†
tHb Capillary 65 20.11 0.25 0.14
tHb tHb 53 20.11 0.08 0.11
tHb Syringe 62 20.27 0.06 0.10
FO
2
Hb Capillary 65 85.19 1.04 0.21
FO
2
Hb tHb 54 86.59 1.43 0.29
FO
2
Hb Syringe 62 84.89 1.17 0.26
FCOHb Capillary 65 5.13 -0.90 0.14
FCOHb tHb 54 6.29 -0.91 0.32
FCOHb Syringe 62 5.13 -0.89 0.16
FMetHb Capillary 65 5.41 -0.20 0.07
FMetHb tHb 54 6.10 -0.39 0.07
FMetHb Syringe 62 5.45 -0.36 0.06
FHHb tHb 54 1.01 -0.75 0.11
FHHb Capillary 65 4.27 -0.36 0.13
Table E-64 Rapidlab 1265 CO-ox Module Recovery and Precision with Total
Hemoglobin of ~14 g/dL and Oxyhemoglobin > 75%
Analyte Mode
*
n
Mean
Reference Mean Bias WRSD
†