Rapidlab 1200 Operator’s Guide: System Specifications E-69
02087462 Rev.V
Micro-
capillary
107 1.011 -0.480 3.939 0.996 6.0 232.5
Micro-
syringe
216 0.991 -0.772 3.258 0.997 5.9 232.3
Table E-68 Rapidlab 1265 Human Whole Blood Method Comparison—pO
2
Mode N Slope Intercept RMSE r
2
Min Max
Capillary 198 0.999 1.888 5.745 0.999 29.2 676.3
Syringe 410 0.979 4.294 8.343 0.998 28.3 683.6
Micro-
capillary
107 0.967 3.922 7.086 0.999 26.3 674.4
Micro-
syringe
216 0.964 3.089 8.702 0.998 26.4 676.1
Table E-69 Rapidlab 1265 Human Whole Blood Method Comparison—Na
+
Mode N Slope Intercept RMSE r2 Min Max
Capillary 193 0.991 1.159 1.506 0.994 107.6 180.2
Syringe 387 1.003 -0.417 1.528 0.995 105.4 182.8
Micro-
capillary
137 1.006 -0.603 1.532 0.995 108.1 189.5
Micro-
syringe
209 1.021 -2.613 1.521 0.995 104.2 187.5
Table E-70 Rapidlab 1265 Human Whole Blood Method Comparison—K
+
Mode N Slope Intercept RMSE r
2
Min Max
Capillary 215 0.987 0.082 0.183 0.997 1.35 12.27
Syringe 431 1.021 -0.084 0.179 0.997 1.02 12.27
Micro-
capillary
143 1.007 0.005 0.182 0.997 1.37 11.85
Micro-
syringe
278 1.015 -0.042 0.169 0.997 1.06 12.27
Table E-71 Rapidlab 1265 Human Whole Blood Method Comparison—Ca
++
Mode N Slope Intercept RMSE r
2
Min Max
Capillary 117 1.088 -0.103 0.092 0.994 0.52 4.59
Syringe 259 1.108 -0.133 0.090 0.995 0.49 4.86
Micro-
capillary
71 1.008 -0.016 0.050 0.998 0.52 4.35
Table E-67 Rapidlab 1265 Human Whole Blood Method Comparison—pCO
2
Mode N Slope Intercept RMSE r
2
Min Max