Rapidlab 1200 Operator’s Guide: System Specifications E-71
02087462 Rev.V
pH, Glu,
Lac
Syringe
218 0.817 0.392 0.494 0.991 0.7 16.1
Table E-75 Rapidlab 1265 Human Whole Blood Method Comparison—tHb
Mode N Slope Intercept RMSE r
2
tHb 296 0.961 0.657 0.515 0.991
Capillary 326 0.963 0.789 0.490 0.992
Syringe 331 0.967 0.620 0.479 0.993
Table E-76 Rapidlab 1265 Human Whole Blood Method Comparison—
FO
2
Hb
Mode N Slope Intercept RMSE r
2
tHb 276 0.999 -0.468 2.188 0.995
Capillary 308 1.002 -0.727 1.740 0.996
Syringe 310 0.995 0.072 1.837 0.996
Table E-77 Rapidlab 1265 Human Whole Blood Method Comparison—
FCOHb
Mode N Slope Intercept RMSE r
2
tHb 193 0.978 0.078 1.095 0.999
Capillary 208 0.994 -0.130 0.921 0.999
Syringe 205 0.990 0.094 0.997 0.998
Table E-78 Rapidlab 1265 Human Whole Blood Method Comparison—
FMetHb
Mode N Slope Intercept RMSE r
2
tHb 208 0.981 0.029 1.019 0.999
Capillary 212 1.025 0.015 0.914 0.999
Syringe 205 0.978 0.027 0.992 0.999
Table E-79 Rapidlab 1265 Human Whole Blood Method Comparison—
FHHb
Mode N Slope Intercept RMSE r
2
tHb 136 1.001 0.535 1.514 0.998
Capillary 144 0.991 0.661 1.554 0.998
Syringe 152 0.954 0.442 1.282 0.999
Table E-74 Rapidlab 1265 Human Whole Blood Method Comparison—Lactate
Mode N Slope Intercept RMSE r
2
Min Max