5-18
Agilent 7500 ICP-MS ChemStation Operators Manual
Creating a Method
5 Edit the equation in the table.
For example, if the interference equation for As is:
75
As =
75
M x 1 +
77
M x (-3.1278) +
78
M x 1.0177
(where
75
M represents 75 AMU), it would be edited.
This equation is taken from EPA Method 6020 (CLP-M Ver.9).
NOTE The edited interference correction equation is a user-defined interference correction
equation to be used only with the currently created method. The default library file
will not be updated. To save a user-defined interference correction equation, click on
Save and enter the name. The equation is saved as a library file. The library file is
added to the drop-down list and can be used for other methods.
6Click OK when you are finished editing for that mass.
ChemStation will return to the Interference Equation dialog box, where you can
select the next mass for which you want to edit the interference equation.
7Click OK after you are finished editing the interference equations.
The Acquisition Mode dialog box appears. To continue creating the method, see
“Setting the Acquisition Mode” in the next section.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals