Agilent 7500 ICP-MS ChemStation Operator’s Manual
Performing a Quantitative Analysis
External Calibration is the most common method of performing quantitative
analysis.
To load an element list from the current method, select the Load Masses from
Current Acq. Params check box.
5Click OK.
The New dialog box will close and a new calibration file will be created.
6 Enter a file name in Calibration Title.
Selecting a Data File
To create a new calibration file, select the data file of a sample standard.
1 Click the Standard Files tab.
The Standard Files tab screen will appear. Select the data file of a sample standard.
Standard Files Tab
2 Select the Reload all masses to calib table check box.
When a data file is chosen with this check box selected, the masses contained in the
data file selected on this tab screen will be loaded to the Table tab screen.
3 In the table on the right side, click on the Data File field of the level to which the
datafile is set.
The data file to be set must correspond to the selected level. For example, when
Level 1 is selected, the data file to be set must represent the calibration solution for
the first level of concentration. The first level is normally a blank solution.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals