EasyManuals Logo

Agilent Technologies 7500 ICP-MS User Manual

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #479 background imageLoading...
Page #479 background image
13-13
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Performing a Semiquantitative Analysis
Loading Factors from a Calibration File
ChemStation enables you to load semiquantitative factors from a calibration file.
1Click Load Factors from Calib.
The SemiQuant Parameters dialog box will appear.
SemiQuant Parameters Dialog Box
2 To load data from a calibration file that is being edited and has not been saved,
click Yes . To load data from an existing calibration file, click No.
Clicking No will bring up the Select Calibration File dialog box.
Select Calibration File dialog box.
3 Select a calibration file using one of the following methods.
• Select a file name from the displayed list and click OK.
• Double-click a file name in the displayed list.
• Enter a file name and click OK.
NOTE If multiple tune steps exist in the selected calibration file, the Select Tune Step dialog
box will appear. Select a tune step using this dialog box.
If multiple isotopes exist in the selected calibration file, the Select Mass dialog box
will appear. Select an isotope using this dialog box.
The Select Calibration File dialog box will close and the selected calibration file
will be loaded.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals