13-13
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Performing a Semiquantitative Analysis
Loading Factors from a Calibration File
ChemStation enables you to load semiquantitative factors from a calibration file.
1Click Load Factors from Calib.
The SemiQuant Parameters dialog box will appear.
SemiQuant Parameters Dialog Box
2 To load data from a calibration file that is being edited and has not been saved,
click Yes . To load data from an existing calibration file, click No.
Clicking No will bring up the Select Calibration File dialog box.
Select Calibration File dialog box.
3 Select a calibration file using one of the following methods.
• Select a file name from the displayed list and click OK.
• Double-click a file name in the displayed list.
• Enter a file name and click OK.
NOTE If multiple tune steps exist in the selected calibration file, the Select Tune Step dialog
box will appear. Select a tune step using this dialog box.
If multiple isotopes exist in the selected calibration file, the Select Mass dialog box
will appear. Select an isotope using this dialog box.
The Select Calibration File dialog box will close and the selected calibration file
will be loaded.