A-19
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Appendix
Quantitation
Internal Standard
When an internal standard is selected in the current calibration curve, the count of
the target ion of the sample data is divided by the ratio of the count per
concentration of the internal standard in the sample data. In this calculation, the
concentration of the internal standard in the first level of the calibration curve is
used as the concentration of the internal standard in the sample data. The count
value for each element reported on a quantitative report is therefore a count ratio.
where
This value is used as the measured value of y in the following sections.
Calculating Concentration
Concentration of the target ion of the sample data is calculated as follows.
where
()
y
y
yx y
y
x
ii
i
===
σσ
// ( if )0
x
y
y
i
i
:
:
:
concentration of the internal standard of the first level in the calibration curve
count of internal standard
count of the target ion
σ
x
y
a
b
:
:
:
c:
concentrat
i
on o
f
t
h
e target
i
on
measured count of the target ion
coefficient "a" in the calibration curve
coefficient "b" in the calibration curve
coefficient "c" in the calibration curve
:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals