Agilent 7500 ICP-MS ChemStation Operator’s Manual
Running a Sample Analysis
Acquiring Sample Data
After loading the method, there are two ways to run a sample manually, Method Run
and Acquired Data. To use Method Run, complete the following steps from the Top
window:
1 Select Methods>> Run.
The Start Run dialog box appears.
Start Run Dialog Box
2 Check the data file directory path in the Data File Name text box and correct the
path if necessary.
3 To use the autosampler, enter a vial number in the Vial text box.
4 Type a sample name and comment.
5 Set a dilution factor if necessary.
Enter a dilution factor directly or click Calculate and follow the procedure
described below.
1) Click Calculate.