Agilent 7500 ICP-MS ChemStation Operator’s Manual
Creating a Method
Isotope Analysis
Isotope Analysis is similar to Spectrum Analysis. The difference is that the dwell
time per mass is 10 times shorter. This mode should be used when precision for iso-
topic information (isotopic ratio and isotope dilution) is required (e.g. radioactive
isotope analysis for determining age, etc.).
To set the Acquisition Mode to Isotope Analysis, complete the following steps:
1 Click the Isotope Analysis button.
The button is marked and the selected mode is outlined.
Acquisition Mode Dialog Box---Isotope Analysis
2Click OK.
ChemStation sets Isotope Analysis as the acquisition mode for the method and the
Isotope Analysis Acquisition Parameters dialog box appears