EasyManuals Logo

Agilent Technologies 7500 ICP-MS User Manual

Agilent Technologies 7500 ICP-MS
659 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #645 background imageLoading...
Page #645 background image
A-29
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Appendix
Appendix C
This appendix includes descriptions for the dead time calibration.
Dead Time Calibration
Dead time is the interval following the arrival of an ion at the detector, during which
the counting of a pulse takes place. It is determined by the width of the multiple
output pulse, electronic pulse widths and circuit recovery times. Another ion
arriving during this time will not be recorded.
A typical value of dead time is 40 nsec
Dead time correction is performed in the following calculation.
N = n / (1-nT)
N:Count after correction (cps)
n: Count before correction (cps)
T: Correction parameter (= Dead Time)
NOTE Dead time calibration needs to be performed after replacement of the electron
multiplier.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals