8-23
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Running a Sample Analysis
Stopping a Run by Stopping a Sequence
If you want to stop the sequence you are currently running, click the Abort push
button from the Run Sequence status window. A dialog box will pop up asking if
you are sure you want to abort. Click Yes , and the entire sequence becomes inactive.
ChemStation will, however, finish acquiring data from any method it has already
begun, unless Stop Run has been pressed in the Real Time Display window.
ChemStation will, however, finish acquiring data from any method it has already
begun, unless Stop Run has been pressed in the Real Time Display window.
If you start the same sequence once again, ChemStation will ask you if you want to
start from where you stopped, or from the beginning. Click Yes if you want to
continue, No if you want to start from the beginning. If you start from the beginning
any previously saved files will be overwritten unless otherwise the data directory is
no changed.
Restarting Sequence Dialog Box
NOTE You can also stop just the current acquisition by double-clicking the ICP-MS Data
Acquisition icon, then clicking the Stop Run push button. This procedure will not
stop the sequence itself; it will stop data acquisition and data analysis for the current
sample, and go on to the next line in the Sequence Log Table.
Stopping a Run by Stopping a Chained Sequence
To stop the chained sequence during the run, click the Abort button from the Edit
Chained Sequence table. ChemStation will finish the chained sequence after running
a sequence it has already begun.
To stop the sequence you are currently running, click the Abort push button from the
Run Sequence status window. A dialog box will pop up asking if you are sure you
want to abort. Click Yes , and the entire sequence becomes inactive. ChemStation
will, however, finish acquiring data from any method it has already begun, unless
Stop Run has been pressed in the Real Time Display window.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7500 ICP-MS and is the answer not in the manual?

Agilent Technologies 7500 ICP-MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7500 ICP-MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals