A-26
Agilent 7500 ICP-MS ChemStation Operator’s Manual
Appendix
SemiQuantitation
The counts used in semiquantitation were the counts of the most abundant point of
the mass specified in the semiquant parameters panel for a given element.
There are some exceptions to the execution of interference check protocol as
follows:
Oxide check is not executed for
72
Ge.
Doubly charged ion check is not executed for
7
Li.
Dimer check is not executed for
7
Li,
24
Mg,
60
Ni,
72
Ge.
Argide check is not executed for
60
Ni,
72
Ge.