19-34
Agilent 7500 ICP-MS ChemStation Operators Manual
Installing the Agilent 7500 ChemStation Software (Windows 2000)
Initialization of a Tune File to the Factory
Default Condition
The following procedure allows initialization of a tune file to the factory default
condition.
1 Select Top window >> Instrument >> Tuning.
2 Select File >> Load Factory Defaults.
The Select Tune Condition dialog box will appear.
7500ce Display 7500cs Display
3 Select the tune file to return to the factory default condition.
The tune files displayed will vary depending on the specific model.
7500a: No tune file is displayed.
7500ce: Standard, H2 gas, He gas
7500cs: Standard, cool plasma, H2 gas, He gas
4Click OK.
The selected tune file will be initialized to the factory default condition.